Metrology: measurement of temperatures
Air humidity measurement Temperature measurement Liquid flow rate measurement
Gas flows rate measurement Pressure measurement Air velocity measurement
CALIBRATION
ACCREDITATION
N°2.1116
TEMPERATURE

Download the description of the laboratory

Calibration methods

Three methods are used according to the field of temperature and the type of sensor:

 1 

Calibration in air inside a climatic chamber

The calibration is performed in the air of a climatic chamber by comparison with a 100 - ohm reference platinum resistance thermometer.

TYPE OF DEVICE IMPLEMENTED RESOURCES
Thermo-hygrometer
Thermo-anemometer
Thermo-recorder
Thermo-enregistreur

Calibration of thermoters
in a climatic chamber.

 2 

Calibration in thermostatically controlled baths

The calibration is performed in an overflowing thermostatically controlled bath, by comparison with a 25-ohm reference platinum resistance thermometer. Depending on the temperature, the liquid used is either alcohol, water or oil. The thermometer probe must be at least 60 mm long for a diameter below or equal to 6 mm, 150 mm long for a diameter over 6 mm.

TYPE OF DEVICE IMPLEMENTED RESOURCES
Platinum Resistance Thermometer 4-wire resistance measurement with a multimeter or display of the measurements chain
Liquid-expansion thermometer Instrumental read degree with the help of a cathetometer
Thermoelectric couple Melting ice bath for the 0°C referenceVoltage measurement
with a multimeter or display of the measurements chain
Measurements chain Display of the measurements chainVoltage or current measurement with a multimeter

 3 


Calibration in a furnace

The calibration is performed in a furnace by comparison with a type S thermocouple. The diameter of the thermocouple to be calibrated must be below 10 mm and its length must be over 270 mm.

TYPE OF DEVICE IMPLEMENTED RESOURCES
Thermocouple Melting ice bath for the 0°C referenceVoltage measurement with a multimeter or display of the measurements chain


Calibration conditions

A calibration comprises at least 3 values of temperature


Calibration fields and uncertainties

Underwater devices

TYPE OF DEVICES FIELD ENLARGED UNCERTAINTY CALIBRATION METHOD
Platinum resistance thermometer TRP (Pt25 ou Pt100) - Measurement chain with Pt100

-80°C
to
+215°C
≥ ±0,03°C
 1 
Industrial platinum resistance thermometer (TRPI) ≥ ±0,06°C
Thermocouple ≥ ±0,10°C
Liquid-expansion thermometer ≥ ±0,08°C
Temperature measurements chain ≥ ±0,06°C
Thermocouple +200°C to +660°C ≥ ±1,5°C
 2 
+660°C to +1 050°C ≥ ±2,2°C
Capteur non immergeable -30°C to
+160°C
≥ ±0,20°C
 3 
Measurement chain with Pt100 +100°C to
+200°C
≥ ±0,15°C
+200°C to
+400°C
≥ ±0,25°C
+400°C to
+600°C
≥ ±0,55°C

On site

TYPE OF DEVICES FIELD ENLARGED UNCERTAINTY CALIBRATION METHOD
Industrial Platinum resistance thermometer (TRPI)
-30°C
to
+215°C
≥ ±0,11°C
 1 
Thermocouple ≥ ±0,15°C
Liquid-expansion thermometer ≥ ±0,14°C
Temperature measurement chain ≥ ±0,11°C
Non-underwater devices 0°C à
+70°C
≥ ±0,20°C
 3 

Non-underwater devices

TYPE OF DEVICES FIELD UNCERTAINTY CALIBRATION METHOD
Non-underwater sensor -30°C to +160°C ±0.2°C to ±0.6°C
 1 
 
CENTRE TECHNIQUE DES INDUSTRIES AÉRAULIQUES ET THERMIQUES